Photoelectric Portable Analyzer PPA–02
Application
The device is used for measurements of optical density of solutions as well as density of liquid colored solutions. Unlike the stationary device type KFK, the measuring procedure is much more simplified.
The analyzer is to operate together with standard quartz cuvette (height — 40 mm, width — 24 mm, optical way length — 5 mm). Maximum operating volume of the cuvette — 2 ml.
The analyzer is to operate together with standard quartz cuvette (height — 40 mm, width — 24 mm, optical way length — 5 mm). Maximum operating volume of the cuvette — 2 ml.
The analyzer has the following technical abilities:
- allows operating with open cuvette section at changing external illumination;
- enables to automatically adjust the sensitivity on blank sample reducing the operation preparation process, increased the accuracy of measurements;
- fulfils the programming processing in digital form quickly evaluating the measuring result on liquid crystal indicator;
- stores in memory up to 40 result processing programs, which enables to register a wide range of substances;
- controlled by 4-key keyboard situated on the front panel of the device.
The main technical characteristics
The range of measured optical density, Б | 0.000 — 1.700 |
Absolute error of optical density measurement in the range 0.030 — 0.800, Б | ± 0.03 |
Absolute error of optical density measurement in the range 0.800 — 1.700, Б | ± 0.15 |
Resolution ability, Б | 0.001 |
Measuring time, sec | 2 |
Continuous operating time, h | 8 |
Power supply voltage of network adapter, V | 220 -33+22 |
Power supply voltage of analyzer, V | 9 |
Power consumption, W | 4 |
Overall dimensions, mm | 195х95х55 |
Weight, kg | 0.35 |